Before starting the temperature adjustment process, close drying chamber lid
and follow below process descriptions:
Enter submenu group
<Adjustment>.
Start <Temperature
adjustment> application by
pressing field with its name.
The moisture analyzer displays
a message box on procedure
start, and data on the process
(step number, current
temperature value, heating
time, burn-in (heating) time).
After 8 minutes, the system
opens a window with an on-
screen numeric keyboard.
Using the on-screen keyboard
insert the current temperature
value of the moisture analyzer,
that is indicated on the
external control thermometer.
Press <Accept> key to proceed
to the following adjustment
step.
The moisture analyzer switches
on the halogen heater and for
another 8 minutes it heats the
drying chamber and
simultaneously displays current
temperature value and other
parameters.