- 44 -
chamber of a moisture analyzer. Before starting temperature adjustment
process, place the temperature calibration kit inside the drying chamber,
following below diagrams:
Step 1.
Remove from the drying chamber the
following:
disposable drying pan
drying pan handle
drying pan bracket
drying chamber shield
Step 2.
Assemble in the drying pan the
components of temperature
adjustment kit:
drying chamber shield
thermometer holder and the
thermometer
Step 3.
After assembling the kit start
temperature adjustment process.
CAUTION:
All the activities must be carried out with great caution not to damage
the measuring mechanism of the moisture analyzer.