n parameter, decimal value specifying working mode number. See OMI
command description for detailed information.
Example
Command:
OMG<CR><LF> give current working mode
Response:
OMG_13_OK<CR><LF> the device currently operates in Statistics mode
Send all implemented commands
Format: PC CR LF
Accessible responses:
PC_A_”Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,
LOGIN,LOGOUT,PC" - command carried out, terminal has sent all
implemented commands.
27.3. Manual printout / Automatic printout
Moisture analyzer MA 3Y series enables generating manual or automatic
printouts.
Manual printout: press key on stabilization of indication
(measurement result),
Automatic printout is generated automatically, in accordance to the settings
of automatic printout (see point 11.6 of this user manual).
The content of a printout depends on setting in menu <Standard printout> -
<Weighing printout template> (see point 16.5 of this user manual).
Format of mass printout:
1
2
3
4 -12
13
14
15
16
17
18
stability
marker
space character mass space unit CR LF
Stability marker
[space] if stable
[?] if unstable
[!] air buoyancy correction mode is enabled
[^] if error of exceeding range to + occurs
[v] if error of exceeding range to occurs
Character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point and right justification
Unit
3 characters with left justification