- 141 -
User log out
Format: LOGOUT CR LF
Accessible responses:
LOGOUT OK CR LF
- command understood, user is logged out
ES CR LF
- command not recognized (error in format)
The result release
Format: SS CR LF
Accessible responses:
SS_OK CR LF - command understood, realization in progress
The command simulates pressing PRINT button on an overlay, according to
settings selected for result release.
Internal adjustment/calibration
Format: IC CR LF
Accessible responses:
IC_A CR LF
IC_D CR LF
- command understood, realization in progress
- calibration completed
IC_A CR LF
IC_E CR LF
- command understood, realization in progress
- time limit of result stabilization exceeded
IC_I CR LF
- command understood but not accessible at this
moment
Lock balance keyboard
Format: K1 CR LF
Accessible responses:
K1_I CR LF
- command understood but not accessible at this
moment
K1_OK CR LF
- command completed
Command locks balance keyboard (motion sensors, touch panel) until it gets
unlocked either by K0 command or by switching the balance off.
Unlock balance keyboard
Format: K0 CR LF
Accessible responses:
K0_I CR LF
- command understood but not accessible at this
moment
K0_OK CR LF
- command completed
Start BEEP signal
The command activates BEEP signal for a specified time limit.