24.5.3. Setting the standard mass by declaring the quantity of a
sample
Procedure:
Enter COUNTING PCS. and then:
While inscribing a sample quantity it should be remembered that the
mass of a single piece should not be lower than 0.1 d and the total
mass of a sample (all pcs) should not be lower than 1 d. An error
appears when this two conditions are not performed.
After inscribing a required sample quantity press and you
will see:
If pieces are to be weight in a container put the container first and tare it.
Then put a load of pieces, wait for the equilibrium (
) confirm it by
pressing
.
Software automatically calculates a single piece mass and goes to
Counting pieces: